SOLDER POT CALIBRATION PROCEDURE FOR PRECISION TEMPERATURE CONTROLLED MODELS
Includes Models: SP-500T, SP-150T, SP-600T, and SP-750T
  1. Turn power on and preset LCD readout to desired temperature and wait 15 to 20 minutes to stabilize.

  2. Insert your calibrated test probe to a depth of 1/2" into the solder at the center point of the crucible and wait 2 minutes for temperature stability. If not the same, proceed with step 3.
     
  3. Remove chassis plug from left side of control box. Insert a plastic screwdriver (DO NOT USE METAL) through hole into slot in potentiometer. Turning clockwise decreases the LCD display, and turning counter-clockwise increases the LCD display

  4. When calibration is complete, re-install the chassis plug into hole.